Mechanical state assessment of the implant-bone continuum: a better understanding of the Periotest method.

  • J. Tricio*
  • , P. Laohapand
  • , D. van Steenberghe
  • , M. Quirynen
  • , I. Naert
  • *Autor correspondiente de este trabajo

Producción científica: Contribución a una revistaArtículorevisión exhaustiva

55 Citas (Scopus)

Resumen

The aim of this study was to obtain a better understanding of the Periotest method when used to detect subclinical mobility of osseointegrated implants. Four hundred two screw-shaped implants were tested with the Periotest device at the time of abutment connection. Several factors, including jaw location, implant and abutment length, and gender, were related to Periotest values (PTVs). Implants located in the anterior region of the mandible showed the lowest mean PTV (-3.2). The influence of abutment and implant length upon PTVs could be detected in the maxilla. In the mandible, only abutment length had influence on PTVs. Women showed higher mean Periotest scores in the maxilla compared with men. This difference was not found in the mandible. The Periotest method, its clinical limitations, and the meaning of a given PTV are also discussed.

Idioma originalInglés
Páginas (desde-hasta)43-49
Número de páginas7
PublicaciónThe International journal of oral & maxillofacial implants
Volumen10
N.º1
EstadoPublicada - ene. 1995
Publicado de forma externa

Huella

Profundice en los temas de investigación de 'Mechanical state assessment of the implant-bone continuum: a better understanding of the Periotest method.'. En conjunto forman una huella única.