Resumen
Objective
Quantify the effect of the osseodensification Densah® protocol on the insertion torque, ISQ, and the removal torque of conical BioHorizons® implants.
Materials and methods
An ex vivo model over fresh pig tibia bone was used. Test group (TG) included 50 osteotomies using Densah® osseodensification protocol, and the control group (CG), 50 osteotomies using BioHorizons®’s recommended procedure. Conical BioHorizons® implants (3.8 × 10.5 mm) were implanted, verifying the insertion torque with a manual torque meter. ISQ values were registered with Ostell® device. Finally, implants were removed with manual reverse torque registering the values. Results were analyzed and compared with the Mann-Whitney test and t-test.
Results
Median and interquartile range per group were as follows: insertion torque, CG: 26 (12) Ncm; TG: 42 (26) Ncm, removal torque, CG: 25 (20) Ncm; TG: 40 (28) Ncm, ISQ value, CG: 69.25 (5.5); TG: 71.5 (4). All variables were significantly higher (p ≤ 0.05) in the osseodensification group.
Conclusions
The Osseodensification technique may improve primary stability in the clinical scenario on tapered implants. Further human RCTs are necessary to validate this.
Quantify the effect of the osseodensification Densah® protocol on the insertion torque, ISQ, and the removal torque of conical BioHorizons® implants.
Materials and methods
An ex vivo model over fresh pig tibia bone was used. Test group (TG) included 50 osteotomies using Densah® osseodensification protocol, and the control group (CG), 50 osteotomies using BioHorizons®’s recommended procedure. Conical BioHorizons® implants (3.8 × 10.5 mm) were implanted, verifying the insertion torque with a manual torque meter. ISQ values were registered with Ostell® device. Finally, implants were removed with manual reverse torque registering the values. Results were analyzed and compared with the Mann-Whitney test and t-test.
Results
Median and interquartile range per group were as follows: insertion torque, CG: 26 (12) Ncm; TG: 42 (26) Ncm, removal torque, CG: 25 (20) Ncm; TG: 40 (28) Ncm, ISQ value, CG: 69.25 (5.5); TG: 71.5 (4). All variables were significantly higher (p ≤ 0.05) in the osseodensification group.
Conclusions
The Osseodensification technique may improve primary stability in the clinical scenario on tapered implants. Further human RCTs are necessary to validate this.
Idioma original | Inglés |
---|---|
Páginas (desde-hasta) | 625-628 |
Número de páginas | 4 |
Publicación | Journal of Oral Biology and Craniofacial Research |
Volumen | 10 |
N.º | 4 |
DOI | |
Estado | Publicada - 1 oct. 2020 |