Abstract
Materials and methods: An ex vivo model over fresh pig tibia bone was used. Test group (TG) included 50 osteotomies using Densah® osseodensification protocol, and the control group (CG), 50 osteotomies using BioHorizons®’s recommended procedure. Conical BioHorizons® implants (3.8 × 10.5 mm) were implanted, verifying the insertion torque with a manual torque meter. ISQ values were registered with Ostell® device. Finally, implants were removed with manual reverse torque registering the values. Results were analyzed and compared with the Mann-Whitney test and t-test. Results: Median and interquartile range per group were as follows: insertion torque, CG: 26 (12) Ncm; TG: 42 (26) Ncm, removal torque, CG: 25 (20) Ncm; TG: 40 (28) Ncm, ISQ value, CG: 69.25 (5.5); TG: 71.5 (4). All variables were significantly higher (p ≤ 0.05) in the osseodensification group. Conclusions: The Osseodensification technique may improve primary stability in the clinical scenario on tapered implants. Further human RCTs are necessary to validate this.
Original language | American English |
---|---|
Pages (from-to) | 625-628 |
Number of pages | 4 |
Journal | Journal of Oral Biology and Craniofacial Research |
Volume | 10 |
Issue number | 4 |
DOIs | |
State | Published - 1 Oct 2020 |
Keywords
- Densah
- Dental implant stability
- Osseodensification